Implantable intraocular pressure monitoring systems: Design considerations
Type
Conference PaperKAUST Department
Computer, Electrical and Mathematical Sciences and Engineering (CEMSE) DivisionElectrical Engineering Program
Integrated Microwave Packaging Antennas and Circuits Technology (IMPACT) Lab
Physical Science and Engineering (PSE) Division
Sensors Lab
Date
2013-12Permanent link to this record
http://hdl.handle.net/10754/564828
Metadata
Show full item recordAbstract
Design considerations and limitations of implantable Intraocular Pressure Monitoring (IOPM) systems are presented in this paper. Detailed comparison with the state of the art is performed to highlight the benefits and challenges of the proposed design. The system-on-chip, presented here, is battery free and harvests energy from incoming RF signals. This low-cost design, in standard CMOS process, does not require any external components or bond wires to function. This paper provides useful insights to the designers of implantable wireless sensors in terms of design choices and associated tradeoffs. © 2013 IEEE.Conference/Event name
2013 IEEE MTT-S International Microwave Workshop Series on RF and Wireless Tecbhnologies for Biomedical and Healthcare Applications, IMWS-BIO 2013ae974a485f413a2113503eed53cd6c53
10.1109/IMWS-BIO.2013.6756252