3-Coil resonance-based wireless power transfer system for implantable electronic

Abstract
This paper presents a 3-coil resonance-based wireless power transfer (R-WPT) system using a single layer of inductor coil windings, in a pancake configuration, in order to obtain a compact system for implantable electronic applications. A theoretical analysis and experimental measurements in terms of quality factor Q and power transfer efficiency (PTE), was done. Our proposed 3-coil scheme can achieve a high PTE with a resonance frequency of 2.46 MHz over a transfer distance of up to 30 mm, by using two 15-mm radius implant coils. The achieved experimental PTE is more than 85%at a 5 mm separation distance, and about 50% PTE at a distance of 20 mm. © 2013 IEEE.

Citation
Yi, Y., Buttner, U., Fan, Y., & Foulds, I. G. (2013). 3-Coil resonance-based wireless power transfer system for implantable electronic. 2013 IEEE Wireless Power Transfer (WPT). doi:10.1109/wpt.2013.6556925

Publisher
Institute of Electrical and Electronics Engineers (IEEE)

Journal
2013 IEEE Wireless Power Transfer (WPT)

Conference/Event Name
2013 IEEE Wireless Power Transfer, WPT 2013

DOI
10.1109/WPT.2013.6556925